Common Issues of Filter Performance Testing Machine
-

1. Machine model not displayed on the home page
Solution: Go to "Machine Model Settings"; confirm the machine model file has been saved; return to the test interface; click "Change Product" to reselect the machine model.

2. Unable to establish instrument connection
Check: Is the instrument powered on; are the communication cables connected; are the port number, serial port number, and IP address correct; is the instrument occupied by other software; are there specific errors in the operation log.

3. Unable to start automatic operation
Check: Has the machine model been selected; is the instrument connection enabled; is the line equipment connected properly; are the robot arm, barcode scanner, and controller in an available state; is the page in a locked or abnormal state.

4. MES shows disconnected
Check: Is the MES service started; are the MES address and port correct; is the network communicating properly; is the firewall allowing the corresponding port; the MES connection information in the software operation log.

5. Barcode scanning failed
Check: Is the barcode clear and complete; is the barcode input method correct; are the barcode prefix, suffix, and length configurations correct; is the barcode scanner connected; is the product position suitable for scanning.

6. Abnormal test results
Check: Is the currently selected machine model correct; is the test process enabled; are the upper and lower limits of the test file correct; is the instrument connection and instrument status normal; is the contact between the product and the fixture good; are there abnormalities in the operation log.

Abnormal Handling and Safety Precautions for the Six-Station Turntable Test System
+

If the instrument or PCB is not connected, check the power supply, cables, serial port configuration, and port occupation, and use the "Reset" button at the top. If abnormal noise, odor, smoke, continuous alarms, or mechanism jamming occurs, immediately shut down the machine or press the emergency stop according to on-site regulations, and then contact maintenance personnel.


Instrument not connected: It is recommended to check the power supply, cables, serial port/network, and click Reset;

PCB connection failed: It is recommended to check the PCB power supply, serial port configuration, and port occupation;

MES not connected: Check the network, address, account, and TOKEN; confirm that the MES desktop tool is running;

No response when scanning: Check the barcode scanner, scanning switch, and barcode rules;

Work order switching failed: Confirm instrument connection, ensure the work order exists, and the file is normal.


Tip: Tests such as withstand voltage and bias current involve electrical hazards; do not open the protective area, touch moving parts, or test terminals while the equipment is running.

Common Faults and Troubleshooting Methods for Power-Frequency Transformer Integrated Test System
+

1. Equipment self-test failed

Possible causes: No equipment is enabled; the COM port number is incorrect or not connected; the baud rate does not match; the handshake request/reply parameters are inconsistent with the equipment protocol.

Suggestion: Go to System Configuration -> Device Configuration to check the enable status and serial port parameters of each device; use the "Verify" button on the card to verify individually.


2. Test conditions not loaded after switching work orders

Possible causes: The machine model file associated with the work order does not exist or failed to load; equipment initialization timed out (the system will prompt a timeout for loading test conditions).

Suggestion: Confirm that the machine model file exists and is in the library; check whether the self-test of the relevant equipment has passed.


3. Withstand voltage / Comprehensive / Power meter alarm timeout

Possible causes: Cylinder position is not in place (related logic is not satisfied); conduction control is not effective or equipment response is abnormal; power meter-related equipment/configuration is not enabled or equipment is unavailable.

Suggestion: Check whether the corresponding stage status shows an alarm; if necessary, use Manual Control to debug the conduction and cylinder status (under safe conditions).


4. Fingerprint function unavailable

Possible causes: The fingerprint module is not enabled or the serial port is unavailable; the connection to the fingerprint recognition device failed.

Suggestion: Enable the fingerprint module device card in the system configuration and configure the serial port parameters correctly; the login window will automatically refresh the fingerprint availability status.

Fixture Maintenance and Troubleshooting
+

1. Fixture Maintenance

Clean the fixture regularly to prevent foreign objects from affecting the test;

Regularly check the fixture wiring for any damage;

Replace the fixture probes regularly.


2. Common Troubleshooting

When it is difficult to test during testing, first check whether the probes are in good contact, whether the probes are damaged, whether the probe wiring is intact, and check whether the air pressure is within the specified range;

If the test value is found to be too different from the actual value, check whether the sample is OK, check whether the sample is aligned properly, and check whether the test data is correct;

If the test value is very large or negative during the test, the product may not be in contact with the probes. Check whether the product is placed properly, check whether the probes are damaged, check whether the fixture is installed correctly, and check whether the test cable is loose.

  • Business
    Phone: 13809631553 (Mr. Zhang)
    Email: ke.zhang@rmeter.cn
  • Technical Support
    Tel: 13688910500 (Mr. Sun)
    Email: chengdan.sun@rmeter.cn
  • After-sales Repair
    Tel: 13713127352 (Mr. Zhou)
    Email: hai.zhou@rmeter.cn
Copyright © RMETER   粤ICP备2026100643号
Copyright © RMETER
粤ICP备2026100643号
Sales
13809631553

After-sales service
13713127352

WeChat

技术咨询